In order to give you a better service Tintometer uses cookies. By continuing to browse the site you are agreeing to our use of cookies.

Pure water conductivity cell for SD 320 with 1 m cable

Mark as favourite

Part Number 19805046

Highlights

  • Precise 2-pole stainless steel electrodes (K ≈ 0.1 cm-1) for the determination of low conductivities (≤ 200 µS/cm, also suitable for ultrapure water)
  • Elimination of influences by atmospheric CO2 via a flow cell suitable for 12 mm probes
  • Automatic temperature compensation (NTC 10 kΩ)

The SD conductivity cell type LC 10 has been developed for conductivities below 200 µS/cm. With the use of two sensitive stainless-steel electrodes even the conductivity of ultrapure water can be determined easily and with highest accuracy. The integrated flow cell made of glass eliminates fluctuations in measured values caused by the introduction of CO2 from the air. The built-in NTC 10 kΩ thermistor provides the necessary automatic temperature compensation.

Operating Temperature -5 - 80 °C
Sensor Material Stainless Steel
Sensor Type 2-Pole
Integrated Temperature Sensor NTC 10 kΩ
Outer Material Stainless Steel
Measuring Range < 200 µS/cm
Cell Constante K ≈ 0,42 cm-1
Length 75 mm
Diameter 12 mm
Cable Length 1.0 m