Part Number 6124683
Turbidity is the key measurement parameter for determining drinking water quality.
To develop this instrument, Lovibond® Tintometer® assembled a team of globally recognized turbidity experts. We tasked them with creating a new process instrument that addresses all of the issues customers struggle with while using their current turbidity systems. These advancements, along with the addition of state-of-the-art communications and user interface make the PTV 1000 and PTV 2000 the next generation of process turbidimeters.
For further information, films and interactive view go to PTV MICROSITE >>
Light Source | LED (Infra red) |
Power Requirements | 90 to 264 VAC, 50/60 Hz. Auto Select |
Accuracy | ± 2 % of reading from 0 to 10 NTU ± 4 % of reading between 10 to 100 NTU |
Stray Light | smaller 0.005 / 5 mNTU |
Displayed Resolution | up to 0.0001 NTU (range dependent) or 5 digits displayed |
Repeatability / Precision | better than 1% at 1 NTU |
Initial Response | 10 % change: 15 seconds @ max flow |
Step Response | T-90 |
Signal Averaging | User Selectable: 1, 3, 6, 10, 30, 60, and 90 seconds, defaulted to 30 seconds |
Sample Temperature | 0 - 50 °C max sample temperature: 70 °C |
Calibration | One Point Calibration at 5.0 or 20 NTU with any regulatory approved formazin |
Sample Flow | 30 to 500 ml/min, optimal flow: 50 to 80 ml/min |
Sample Inlet Connection | 1/4-inch NPT female, 1/4-inch compression fitting tubing (included) |
Sample Outlet (Drain) Connection | 3/8-inch NPT female, 3/8-inch hose barb tubing (included) |
Sample Inlet Tubing | 1/4-inch OD or 6 mm OD |
Sample Oulet Tubing | 3/8-inch OD or 9 mm OD |
Turbidimeter Body Drain | Quick connect with integrated check valve |
Analog Output: Measurement Module | 1 selectable 0 - 20 mA or 4 - 20 mA; output span programmable over any portion of the measurement range |
Analog Output: Junction Box | 1 selectable 0 - 20 mA or 4 - 20 mA; output span programmable over any portion of the measurement range |
Alarms | Three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated 5 A resistive load at 230 V, AC |
Enclosure Type: Junction Box | Fiber reinforced polyester |
Enclosure Rating: Junction Box | IP 66 |
Measuring Range | 0.0001 - 100 NTU |
Safety | Listed by TÜV Rheinland to UL 61010A-1: Certified by TÜV Rheinland to CSAC22.2 No. 1010.1: CE Certified by TÜV Rheinland to EN 61010-1 |
Immunity | CE certified by TÜV Rheinland to EN 61326 (Industrial Levels) |
Emissions | Class A: EN 61326, CISPR 11, FCC Part 15, Canadian Interference-Causing Equipment Regulation ICES-003 |
Mounting Hardware | Turbidimeter Sensor - Slotted Mounting Bracket that can be affixed to any vertical surface or panel (Optional). Junction Box - Direct mounting to any vertical surface or panel (Optional) |
Method of Verification | Wet Standards or dry verification device |
Limit of Determination | 0 NTU |
Limit of Detection | 0 NTU |
Flow Indicator | true |
Portability | Process |
Environmental Conditions | 5 - 50 °C, rel. humidity 5 - 95 % (Non-condensing) |
Compliance | ISO 7027 |
Protection Class | IP 65 |
Dimensions | 158.4 x 340.4 x 334.5 mm |
Weight | 5 kg |
Title | Part Number | Mark as favourite |
---|---|---|
T-CAL-Standard, 0.30 NTU, Standard for Verification | 48010035 |
|
T-CAL-Standard, 1.00 NTU, Standard for Verification | 48010135 |
|
T-CAL-Standard, 20.00 NTU, Standard for Calibration | 48010335 |
|
T-CAL-Standard, 5.00 NTU, Standard for Calibration | 48010235 |
|
T-CAL-Standard, 0.30 NTU, 500 ml | 48011050 |
|
T-CAL-Standard, 1.00 NTU, 500 ml | 48011150 |
|
T-CAL-Standard, 20.00 NTU, 500 ml | 48012350 |
|
T-CAL-Standard, 5.00 NTU, 500 ml | 48012250 |
|
Calibration Tube Replacement (Blue) | 19806-062 |
|
Cleaning Tube Assembly (Black) | 19806-072 |
|
Scale Remover | 54013003 |
|
Detergent Solution | 54011010 |
|
Complete Cleaning Kit | 19806-063 |
|
Cleaning Solution | 54010435 |
|
Cleaning Accessory Kit | 19806-112 |
|